‹ All models

Invacare Fox – fault codes, parts & serial label

Typepowerchair (rear-wheel drive)
ControllerInvacare LiNX (LiNX builds listed in the LiNX service manual) INFERRED
Manualinvacaredocs.com/assets/documents/PIM/User_Manual_Linx_(en).pdf ↗

Fault codes (8)

From the LiNX powerchair control system (Invacare LiNX) code set. All LiNX powerchair control system (Invacare LiNX) codes

CodeMeaningChecks
1 flashRemote faultSwitch off and on; check the remote cable; contact the provider
2 flashesNetwork / configuration faultCheck all cables, recharge the batteries, check the charger
3 flashesMotor 1 faultCheck the motor plug and wiring
4 flashesMotor 2 faultCheck the motor plug and wiring
5 flashesLeft brake faultPut the freewheel lever in drive; check the brake plug
6 flashesRight brake faultPut the freewheel lever in drive; check the brake plug
7 flashesModule faultReverse away from any obstacle; switch off and on; contact the provider
All LEDs flashJoystick not centred (drive out of neutral)Release the joystick, switch off and on again

Serial code label

No serial location data yet for this powerchair. Serial codes

Parts (15)

PartPart no.
Castor 200x50mm PU, black rimSP1578146
Wheel 12.5" x 2.25", puncture proof, black plastic rimSP1582006
Wheel 12.5" x 2.25", puncture proof, black plastic rimSP1073935
Rear Wheel 12.5" x 2.25", puncture proof, Aluminium Rim silverSP1576191
Tire 12.5" x 2.25" Puncture Proof23009262
Motor cpl. 6km/h LH SharkSP1574377
Motor cpl. 6km/h RH SharkSP1574378
Motor cpl. 6km/h LH VR2SP1574379
Motor cpl. 6km/h RH VR2SP1574380
Powermodule, Shark, 75A, drive only, DK-PMB011433332
Remote DK-REM D01, Shark2 (Drive only)1492567
Powermodule, VR2, 60A, only drive, VR2 60SP1504056
VR2 Remote (Drive only)SP1504057
Battery, 12V/50Ah C20, AGMSP1522107
Battery, 12V/34Ah C20, AGMSP1521459

Summarised from public manuals and supplier listings. Always check the product's own manual before any repair.