Drive DeVilbiss Enigma Energi+ – fault codes, parts & serial label
Typepowerchair
ControllerPG Drives VR2 (Drive Medical flash-code sheet) CONFIRMED
Fault codes (17)
From the VR2 powerchair controller code set. All VR2 powerchair controller codes
| Code | Meaning | Checks |
|---|---|---|
| 1 bar | Low battery voltage | Check the connections to the battery. If the connections are good, try charging the battery |
| 2 bars | Left motor disconnected | Check all the connections and leads between the left motor and the power module |
| 4 bars | Right motor disconnected | Check all the connections and leads between the right motor and the power module |
| 3 bars | Left motor wiring trip | Check all the connections and leads between the left motor and the battery |
| 5 bars | Right motor wiring trip | Check all the connections and leads between the right motor and the battery |
| 6 bars | Charger connected (drive inhibited) | Disconnect the battery charger; otherwise find the external inhibit signal for this chair |
| 7 bars | Possible joystick trip | Make sure the joystick is in the centre position before switching on the VR2 |
| 8 bars | Possible control system trip | Make sure that all connections are secure |
| 9 bars | Solenoid brake trip | Check the parking brake and motor connections. Make sure all the system connections are secure |
| 10 bars | High battery voltage | Check the battery connections |
| Speed LEDs flash with the gauge (+S) | Communication fault | Check the cable and plugs between joystick and power module; cycle the power |
| Actuator LED flashes with the gauge (+A) | Actuator trip | Check the actuator wiring and plugs |
| Battery gauge steps up | Charging (drive inhibited) | No drive until the charger is unplugged and the power switched off and on |
| Speed indicator ripples up and down | Chair locked | Unlock it with the key or the joystick unlock sequence in the chair manual |
| Speed indicator flashes | Speed limited | Lower the seat to drive at full speed |
| Battery gauge ripples at power-on | Joystick displaced | Centre the joystick, switch off and on again |
| Gauge blinks every 2.5 seconds | Sleep mode | Switch off and on again |
Serial code label
No serial location data yet for this powerchair. Serial codes
Parts (15)
| Part | Part no. |
|---|---|
| LHS motor | Z19464-01 |
| RHS motor | Z19474-01 |
| LHS brake unit | Z19700-01 |
| RHS brake unit | Z19750-01 |
| VR2 joystick | C10-028-00100 |
| VR2 module | C10-028-00200 |
| VR2 dual control and cable (PGD) | Z21071 |
| Battery box pair, VR2 connectors | Z19960-01 |
| Castor wheel ass'y | Z19405-01 |
| Height adj. anti tip wheels | Z19800-01 |
| VR2 joystick bracket | Z19931-01 |
| VSI attendant module bracket | Z21065 |
| RHS handle for VR2 | Z21076PG |
| Centre pin ass'y Energi Plus | Z19010-01 |
| Drive Energi Powerchair Motor Carbon Brush |
Summarised from public manuals and supplier listings. Always check the product's own manual before any repair.