Fault codes (17)

From the VR2 powerchair controller code set. All VR2 powerchair controller codes

CodeMeaningChecks
1 barLow battery voltageCheck the connections to the battery. If the connections are good, try charging the battery
2 barsLeft motor disconnectedCheck all the connections and leads between the left motor and the power module
4 barsRight motor disconnectedCheck all the connections and leads between the right motor and the power module
3 barsLeft motor wiring tripCheck all the connections and leads between the left motor and the battery
5 barsRight motor wiring tripCheck all the connections and leads between the right motor and the battery
6 barsCharger connected (drive inhibited)Disconnect the battery charger; otherwise find the external inhibit signal for this chair
7 barsPossible joystick tripMake sure the joystick is in the centre position before switching on the VR2
8 barsPossible control system tripMake sure that all connections are secure
9 barsSolenoid brake tripCheck the parking brake and motor connections. Make sure all the system connections are secure
10 barsHigh battery voltageCheck the battery connections
Speed LEDs flash with the gauge (+S)Communication faultCheck the cable and plugs between joystick and power module; cycle the power
Actuator LED flashes with the gauge (+A)Actuator tripCheck the actuator wiring and plugs
Battery gauge steps upCharging (drive inhibited)No drive until the charger is unplugged and the power switched off and on
Speed indicator ripples up and downChair lockedUnlock it with the key or the joystick unlock sequence in the chair manual
Speed indicator flashesSpeed limitedLower the seat to drive at full speed
Battery gauge ripples at power-onJoystick displacedCentre the joystick, switch off and on again
Gauge blinks every 2.5 secondsSleep modeSwitch off and on again