Fault codes (14)
From the Elite SRE-2010 – under-seat diagnostic codes code set. All Elite SRE-2010 – under-seat diagnostic codes codes
| Code | Meaning | Checks |
|---|---|---|
| A2 | Seat folded up | Lower the seat |
| A3 | Seat out of riding position | Turn seat to lock |
| A7 | Seat or footrest jammed | Clear obstruction; dealer |
| C1 | Charging | Nothing to do |
| C2 | Off charge (running) | Nothing to do |
| C4 | Parked, charger not detected | Check outlet and charge contacts |
| C5 | Battery critical | Charge immediately |
| C6 | Battery low | Charge |
| L1 | Seat safety switch open | Check seat position; dealer |
| L2 | Up limit open | Normal at top; Dealer otherwise |
| L3 | Down limit open | Normal at bottom; Dealer otherwise |
| L6 | Mid-rail park active | Nothing to do |
| L8 | Soft stop switch | Nothing to do |
| Beeps at top, seat turned | Multi-user warning | Lock the seat forward (dealer can set single-user) |